電話:
021-31265669傳真:
021-51861312
沃辛頓LS6000有一款入門級(jí)自動(dòng)控制的液氮供給系統(tǒng)LS6000AT,沃辛頓液氮罐在2016年后推出一款LS6000-CS,在此我們對(duì)比兩款的異同點(diǎn)。
控制功能項(xiàng) | CS100控制器 | AUTO-TEND |
產(chǎn)品圖片 | ![]() | ![]() |
控制面板 | 有 | 有 |
自動(dòng)灌充 | 有 | 有 |
灌充開關(guān)顯示 | 有 | 有 |
高溫報(bào)警 | 有 | |
低溫報(bào)警 | 有 | 有 |
遠(yuǎn)程報(bào)警 | 有 | |
傳感器感應(yīng)失敗報(bào)警 | 有 | |
報(bào)警失效提示 | 有 | |
**遙控調(diào)試 | 有 | |
液氮供給中斷報(bào)警 | 有 | |
加電自檢報(bào)警 | 有 | |
熱電偶校準(zhǔn)報(bào)警 | 有 | |
開蓋時(shí)間過(guò)長(zhǎng)報(bào)警 | 有 | |
2通道網(wǎng)絡(luò)線數(shù)據(jù)接口 | 有 | |
接電腦/記錄/打印 | 有 | |
溫度控制 | 有 | |
數(shù)據(jù)自動(dòng)儲(chǔ)存記錄 | 有 | |
手動(dòng)除霧 | 有 | |
自動(dòng)除霧 | 有 |
為了確保LS系列的正確運(yùn)行和保持優(yōu)良性能,CS100應(yīng)遵循常規(guī)的維護(hù)計(jì)劃。這將包括以下內(nèi)容:
Display Brightness LS SERIES V0.5 Firmware SER 12345 Serial Number
To ensure proper operation and maintain excellent performance of the LS SERIES, a regular maintenance schedule should be followed for the CS100. This would include the following:
CS100控制系統(tǒng)
檢查是否暴露于濕氣、磨損、連接器問(wèn)題和面板或按鈕損壞。此外,定期固件更新可能是重要的。
CS100 CONTROL SYSTEM
Examine for exposure to moisture, wear and tear, connector problems, and damage to the faceplate or buttons. In addition, periodic firmware updates may be important.
電磁閥
檢查電線和連接器是否損壞。每2年更換一次。液位傳感器總成,檢查電線和連接器的損壞情況。更換每2年熱電偶組件檢查損壞電線和連接器。
Solenoid Valve
Examine wires and connector for damage. Replace every 2 years. Level Sensor Assembly
Examine for damage to wires and connector. Replace every 2 years Thermocouple Assembly Examine for damage to wires and connector
電源
對(duì)電源和電源線損壞的檢查
此外,還應(yīng)對(duì)冷凍機(jī)及其機(jī)械部件進(jìn)行檢查和預(yù)防性維護(hù)。詳情請(qǐng)參閱業(yè)主手冊(cè)。
Power Supply
Examine for damage to power supply and power cords
In addition, inspection and preventive maintenance should also be performed on the freezer and its mechanical parts. Refer to owner’s manual for details.
如果觀察到或懷疑冷凍系統(tǒng)的任何間歇操作,即使在正常維護(hù)計(jì)劃之外,也應(yīng)立即進(jìn)行調(diào)查和補(bǔ)救。
如果LS系列液氮罐出現(xiàn)問(wèn)題或出現(xiàn)效率不高的情況,請(qǐng)供應(yīng)商尋求幫助。CS100已納入*診斷工具,以協(xié)助識(shí)別和糾正可能出現(xiàn)的任何問(wèn)題。
If any intermittent operation with the Cryostorage System is observed or suspected, it should be investigated and remedied immediay even if this falls outside of the normal maintenance schedule。
If the LS SERIES freezer experiences problems or appears that it is not operating at optimum efficiency, please contact your supplier for assistance. The CS100 has incorporated state of the art diagnostic tools to assist in the identification and correction of any issues that may arise。